The Role of a Novel Digital Microcapillary Instrument in Detecting Blood and Plasma Hyperviscosity
Abstract: to test the
precision and accuracy of a Digital Microcapillary instrument in measuring
blood and plasma viscosity. Methods: about 40 blood samples were drawn for
precision test. The samples were obtained from patients admitted for Medical
Check-Up at CiptoMangunkusumo Hospital (RSCM) in December 2011. Accuracy test was
evaluated using cross-sectional design and involving 135 patients with acute
ischemic stroke. The patients underwent either inpatients or outpatients care
at Department of Clinical Pathology, Department of Neurology,and Emergency Unit
of RSCM, Fatmawati Hospital Jakarta, Prikasih Hospital Jakarta, and Bhakti Yuda
Hospital Depok. The precision test was evaluated by calculating the coefficient
of variation (CV), interrater variability of Cronbach Alpha, and reliability
coefficient of Bland Altman. The accuracy of the test was evaluated with a
diagnostic test. The gold standard used for these tests was Brookfield LV-DV
III viscometer. Results: the results of precision test were: CV = 0.04;
interrater variability of blood and plasma viscosity = 0.94 and 0.82, respectively;
the Bland Altman mean difference = –0.19. The results of accuracy test were:
sensitivity of blood and plasma viscosity measurement were 88.9% and 100%,
respectively; specificity of blood and plasma viscosity measurement were 88.9%
and 84%, respectively. Conclusion: the digital microcapillary has high
sensitivity and specificity; therefore the instrument can be considered to be
used as screening test tool to measure blood and plasma viscosity.
Key words: digital
microcapillary, blood viscosity, plasma viscosity, acute ischemic stroke
Author: Al Rasyid, Jusuf
Misbach, Jan S Purba, Ina S Timan, Lugyanti Sukrisman, Muchtaruddin Mansyur,
Efta Yudiarsah, Suroto Suroto
Journal Code: jpkedokterangg140325